ISO 14155:2026 What’s Changed and What It Means for Your Clinical Investigation

The fourth edition of the international GCP standard for medical device clinical investigations is now in effect, with no transition period. Here is what sponsors, investigators, and CROs need to know.

A Standard That Has Been a Long Time Coming

ISO 14155 has been the GCP reference for medical device clinical investigations since its first edition in 2003. The third edition, published in July 2020, made significant strides in aligning the standard with the EU Medical Devices Regulation (MDR 2017/745) and the ICH E6(R2) framework. But five years of practical implementation exposed gaps, particularly around risk management, study oversight structures, and statistical planning.

The fourth edition directly addresses those gaps. It introduces more prescriptive requirements, clearer distinctions between types of risk, new formal oversight structures, and a statistical annex that brings medical device investigations much closer to the rigour expected in pharmaceutical trials.

⚠ No transition period, requirements apply immediately

One timing point worth noting: ISO 14155:2020 had only just achieved formal harmonisation with the EU MDR when the fourth edition arrived, an unusual situation that has created legitimate questions for sponsors about the harmonisation status of the new standard. As of publication, EN ISO 14155:2020 remains the harmonised standard for MDR conformity purposes. The harmonisation process for ISO 14155:2026 is ongoing. Sponsors should monitor the EU Official Journal for updates.

The Key Changes: What’s New in ISO 14155:2026

1. Risk Management: A Clearer, More Structured Framework

If applying ISO 14971 risk management principles to clinical investigations has been a grey area in your organisation, the fourth edition addresses this directly.

The key distinction introduced is the separation of two categories of risk:

Device-related risks: including residual risks relevant to the specific study population, sample size, and indication, must be evaluated using full ISO 14971 methodology. The standard recognises that adverse device effect rates during an investigation can differ from post-market use. A small trial population amplifies ethical impact; risks must be contextualised accordingly.

Procedure-related risks: arising from non-routine clinical procedures required by the Clinical Investigation Plan (CIP) but outside standard clinical practice, require a descriptive risk assessment.

A risk based monitoring approach should be defined to manage the oversight of risks associated with the clinical investigation.

The standard clarifies how general device risk management principles translate into study-specific considerations, with focus on investigational device risks, especially residual risks relevant to the study population, sample size, and indication. These updates will require greater rigour when conducting and documenting risk assessments, but the result will be sponsors having more robust, better-documented risk management files tailored to each investigation.

Kirsty Macleod, Head of Clinical Research, MDx CRO

2. Data Monitoring Committees and Clinical Events Committees: Formal Requirements for Both

Study oversight structures receive significant attention in the fourth edition.

For DMCs, ISO 14155:2026 provides further clarification on this requirement. Where a DMC is determined to be in place, sponsors must now pre-define, and document in the CIP, the specific conditions under which a study would be suspended or stopped.

CECs are introduced for the first time in ISO 14155:2026. A Clinical Events Committee is an independent group of clinical experts established by the sponsor to ensure consistent classification and assessment of clinical events across sites. In multi-centre studies, different investigators may characterise the same event type differently, a CEC reduces that variability and strengthens data reliability. Where a CEC is used, its independence, role, and conflict-of-interest management must be formally defined at the outset of the study.

3.Informed Consent: Strengthened Expectations and Ethics Committee Implications

The fourth edition introduces more operational requirements around ethics, subject rights, and consent.

Specific updates include a prohibition on deviations from eligibility criteria without a formal CIP amendment; clearer requirements for explaining future use of health data to subjects; and expanded protection requirements for vulnerable populations and cross-border studies.

Ethics committees will scrutinise these more closely. Sponsors referencing ISO 14155:2026 compliance in submissions should expect greater attention to consent documentation, the consent process itself, and how protocol deviations are governed.

Ethics committees will have altered expectations when sponsors state compliance with the revised standard. Any mismatches between the risk management file and the informed consent form, downplaying known residual risks, using generic language, are routinely flagged during ethics review. The risk file and the consent must tell the same story, in plain language.

Kirsty Macleod, Head of Clinical Research, MDx CRO

4. Design Considerations: The Estimand Framework Arrives

The introduction of Annex K, a new informative annex on clinical investigation design considerations, brings the device world significantly closer to the statistical expectations long standard in pharmaceutical development.

In practice, this means study protocols are expected to more precisely define the clinical question being answered, including how intercurrent events (patient dropouts, treatment switches, protocol deviations) are handled in the primary analysis. Missing data strategies must be planned prospectively, not fixed post-hoc.

For organisations working across both medical devices and pharmaceuticals, this alignment reduces friction. For teams accustomed to simpler device trial designs, it will require additional investment in biostatistics at the planning stage.

5. Clinical Performance: A Sharper Definition With Real Consequences

The definition of “clinical performance” has been updated to more explicitly link device performance to measurable clinical benefit resulting from technical or functional characteristics. This is not a minor editorial change. It has direct implications for how endpoints are selected and justified in the CIP, and how clinical evidence is structured in Clinical Evaluation Reports (CERs).

Sponsors designing new studies need to ensure their primary endpoints map directly to the updated definition. For CER authors, it adds another layer of alignment to consider when assessing clinical data generated under different versions of the standard.

The Most Common Mistakes Sponsors Make regarding ISO 14155, and How to Avoid Them

When running and overseeing clinical investigations, the same errors appear repeatedly when sponsors try to apply risk management requirements to active studies. The fourth edition makes some of these errors significantly harder to ignore during regulatory or ethics review.

The mistakeWhat is expected instead
Reusing the post-market risk file unchanged

Clinical investigations have different risk profiles, smaller populations, learning-curve users, investigational configurations. The ISO 14971 product risk file created for design verification does not translate directly.
A study-specific risk assessment clearly linked to the CIP, subject population, site capabilities, and operator experience. Justified differences between clinical investigation risk and intended-use risk.
Applying full ISO 14971 to procedure risks

Sponsors apply the full methodology to blood draws, biopsies, imaging, and extra clinic visits, logging these as device risks. Regulators increasingly challenge over-engineered risk files because they obscure actual device risk.
Clear separation in documentation: device-related risks, ISO 14971 process; procedure-related risks, descriptive assessment. Explicit statement in the CIP explaining this distinction.
Freezing the risk file once the study starts

Risk management is treated as a design artefact with no named responsibility during execution. Adverse events, near-misses, and protocol deviations that reveal new hazards are not fed back into risk assessment.
Defined triggers for risk review (SADEs, DMC recommendations, deviation trends). Documented updates showing how clinical findings feed back into hazard identification, risk estimation, and benefit-risk evaluation.
Weak linkage between risk file and consent form

The informed consent form downplays known residual risks, uses generic language, or is inconsistent with what appears in the risk management file. Ethics committees view this as a red flag.
Clear mapping between residual risks and subject-facing explanations. Justification where technical risks are simplified or grouped. Immediate consent updates if risk understanding changes.
Risk controls that exist on paper but not in the study

Controls are listed in the risk file but not implemented, training listed but not documented, IFU warnings not reflected in consent, monitoring controls not built into the protocol.
Traceability from risk, control, CIP section, training, monitoring. Cross-checks during monitoring that controls are actively applied. This is one of the most common audit findings.

What Happens to Studies Already Underway?

This is a frequently asked question by sponsors where the answer requires some nuance.

Studies initiated and approved under ISO 14155:2020 remain valid. The publication of the fourth edition does not retroactively invalidate work done in compliance with the previous version. A study designed, approved, and conducted in line with ISO 14155:2020 remains compliant with that version, and that data remains usable to support CE marking, clinical evaluation updates, and post-market evidence, provided it was compliant at the time of initiation.

Where things become more complex is with substantial amendments. If an ongoing study undergoes a significant change after March 2026, a protocol amendment, addition of new sites, introduction of a DMC or CEC, changes to the subject population, regulators and notified bodies may expect the affected areas to be brought into alignment with ISO 14155:2026. This is not a hard rule, but it is a risk-based expectation that should be assessed and documented.

In practice, authorities accept a clear, documented rationale: the clinical investigation was designed and initiated in compliance with ISO 14155:2020, which represented the state of the art at the time of study initiation. The publication of ISO 14155:2026 has been reviewed, and no changes were required that would materially impact subject safety, data integrity, or scientific validity. This approach is widely accepted, provided it is documented, risk-based, and justified.

Kirsty Macleod, Head of Clinical Research, MDx CRO

What Sponsors and Investigators Must Do Now

Given that there is no transition period, the question is not whether to act but how to prioritise. Here is a practical framework:

1. Run a formal gap analysis

Assess your current SOPs, CIP templates, risk management processes, and oversight structures against ISO 14155:2026. Focus first on risk management (device vs. procedure risk distinction), DMC/CEC governance, and statistical planning documentation.

2. Update your QMS and SOPs

Procedures, templates, and SOPs that reference ISO 14155:2020 need revision, particularly around CEC governance (new), DMC governance (strengthened), risk management integration, and informed consent processes.

3. Assess ongoing studies

Review active protocols or protocols in development. For studies with upcoming substantial amendments, prepare a documented assessment of whether and how ISO 14155:2026 requirements apply. Always document the rationale for decisions made, or not made.

4. Deliver structured training

Sharing the updated standard document is not sufficient. The risk management changes in particular require dedicated training for clinical affairs, regulatory affairs, and quality teams, not just a circulated PDF.

5. Monitor the harmonisation process

Keep an eye on the EU Official Journal for updates on the formal harmonisation of ISO 14155:2026 under the MDR. Until harmonised, EN ISO 14155:2020 remains the reference for MDR conformity presumption.

Frequently Asked Questions about ISO 14155:2026

When did ISO 14155:2026 come into effect?

ISO 14155:2026 (Edition 4) was published in March 2026 and replaced ISO 14155:2020 immediately, with no defined transition period. Any new clinical investigation initiated after publication is expected to reference ISO 14155:2026 or provide a clear justification for non-alignment.

Do studies already underway under ISO 14155:2020 need to be updated?

Studies initiated and approved under ISO 14155:2020 remain valid. However, any substantial amendment to an ongoing study after March 2026 may trigger an expectation to align affected parts with ISO 14155:2026, or provide a documented justification explaining why partial or full transition is not appropriate. Always document the rationale.

Is ISO 14155:2026 harmonised with the EU MDR?

Not yet. As of April 2026, EN ISO 14155:2020 remains the harmonised standard providing presumption of conformity with the EU MDR. The formal harmonisation process for ISO 14155:2026 is underway. Sponsors should monitor the EU Official Journal regularly for updates.

What is the difference between a DMC and a CEC under ISO 14155:2026?

A Data Monitoring Committee (DMC) oversees overall trial safety and has the authority to recommend stopping or modifying a study. ISO 14155:2026 requires sponsors to pre-define stopping conditions and justify the absence of a DMC. A Clinical Events Committee (CEC) newly introduced in the 2026 edition, is an independent panel of clinical experts that ensures consistent classification of events across sites in multi-centre studies, improving data reliability.

Does ISO 14155:2026 apply to PMCF studies?

Yes. ISO 14155 applies to post-market clinical investigations, including PMCF studies. Annex I of the standard defines the applicability of requirements to different types of post-market investigations. Some requirements may be modified or exempt for observational studies, but the core GCP principles, including risk management, monitoring, and data governance, apply.

What is the estimand framework introduced in Annex K?

The estimand framework (from ICH E9(R1)) provides a structured approach to defining exactly what clinical question the investigation is designed to answer, and how events that disrupt the intended treatment (dropouts, protocol deviations, treatment switches) should be handled in the primary analysis. It requires these decisions to be made and documented at the design stage, not resolved post hoc in statistical analysis.

Does ISO 14155:2026 apply to IVD clinical performance studies?

ISO 14155 applies to medical devices, not in vitro diagnostics (IVDs). IVD clinical performance studies are governed by ISO 20916:2019. However, where a medical device and an IVD are used in an integrated system, elements of both standards may apply. Sponsors of combination investigations should assess both standards.

    

Related Articles

ISO 14971 Applied to Clinical Investigations: A Practical Guide

PMCF Under MDR: What ISO 14155 Requires

Clinical Investigation Plan: What You Must Include (2026 Update)

EN ISO 14155:2020/A11:2024 and MDR Harmonisation

Written by:
Kirsty Macleod

Kirsty Macleod

Head of Clinical Research

Clinical Operations leader with IVD and CDx device expertise, delivering global clinical studies while establishing best practices and high performing teams.
Industry Insights & Regulatory Updates

EU MDR 2017/745 for Dental Devices: Complete Guide (2026)

Dental device regulation under EU MDR 2017/745 has increased the level of scrutiny applied to manufacturers, particularly in three areas: classification decisions, clinical evidence, and Notified Body (NB) review readiness. The practical consequence is simple: submissions that lack internal consistency generate more questions, longer review cycles, and avoidable delays.

In the 2026 webinar delivered by regulatory experts from MDx CRO, the discussion focused on the most common challenges and the most reliable strategies for dental device manufacturers. The insights were grounded in real-world regulatory work supporting more than 500 CE-marked devices, with an emphasis on reducing regulatory risk and building submissions that remain defensible under NB assessment.

Manufacturer priorities under EU MDR: what drives timelines and outcomes

EU MDR does not typically create delays because manufacturers do not have documents. Delays occur when documentation does not form a single, defensible position. Notified Bodies escalate questions when claims, classification rationale, risk controls, clinical evidence, and post-market plans do not support the same intended purpose and performance narrative.

Key priorities for dental device manufacturers under EU MDR:

  • Establish and maintain a consistent intended purpose and claim set across the file (labeling, IFU, CER, and technical documentation).
  • Build a classification rationale that is explicit, referenced, and difficult to misinterpret.
  • Demonstrate clinical evidence proportionality, clearly linked to risk, novelty, and claims.
  • Prepare for NB review with a coherent evidence package, including traceability from hazards to controls to verification.
  • Ensure PMS and PMCF are designed to confirm performance and safety in real use, not to satisfy a formal requirement.

A submission is reviewed as an argument supported by evidence. When the argument changes between sections of the file, the NB must resolve the conflict through questions. That process is where time is lost.

Step 1. Classify your dental device correctly (Annex VIII)

Classification under Annex VIII is the point where regulatory strategy becomes operational. A small change in intended purpose or in device characteristics can shift the class and materially change clinical evidence expectations, NB involvement, and post-market obligations.

In the webinar, the discussion repeatedly returned to the importance of early discipline around intended purpose. Dental manufacturers often broaden claims for commercial reasons (for example, broad “compatibility” claims or biologically ambitious performance language). Under MDR, that approach frequently increases the burden of proof.

Practical classification anchors in dental:

  • Implants and implantable devices frequently fall under Rule 8, with class outcomes dependent on specific characteristics and risk context.
  • Resorbable materials often face elevated scrutiny because performance and safety evolve as the product degrades and interacts with tissue.
  • Custom-made versus patient-matched frequently causes misclassification and incorrect pathway selection, particularly in digital workflows.
  • Software classification under Rule 11 can change significantly based on clinical impact and decision influence.

A defensible classification process typically answers these questions explicitly:

  • What is the intended purpose and what claims are being made?
  • Is the device invasive, and if so, what level of invasiveness applies?
  • What is the duration of use and the relevant contact type?
  • Is the device implantable, resorbable, or otherwise associated with long-term biological interaction?
  • Does the device incorporate a medicinal substance with an ancillary action?
  • Is any element of the product regulated as software influencing clinical decisions?

Classification is assessed as a justification, not a label. A strong justification reduces NB discretion and stabilises the remainder of the submission.

Step 2. Conformity assessment route by class, and how Notified Bodies evaluate risk

Manufacturers often plan around the conformity assessment route, but Notified Bodies allocate effort based on what drives residual risk and uncertainty. Under MDR, two devices of the same class can attract different levels of scrutiny when novelty, claim strength, or evidence quality differs.

In dental, common drivers of NB scrutiny include:

  • Novel materials and surface treatments particularly where particulate generation, chemical residues, coating stability, or long-term performance need stronger justification.
  • Compatibility and system claims, especially for implant components and abutments, where broad claims are difficult to support without precise system definition and evidence.
  • Borderline product categorisation, such as semi-finished CAD/CAM materials, where intended purpose determines medical device status.
  • Digital workflows, where the manufacturer controls design parameters and outputs are patient-specific within defined constraints.

From a submission strategy perspective, the objective is to reduce uncertainty for the reviewer. That is achieved by ensuring the file communicates a consistent position:

  • The intended purpose and claims are consistent and reflected in labeling and IFU.
  • The classification rationale is explicit and mapped to Annex VIII logic.
  • Risk controls are linked to verification and validation evidence.
  • Clinical evaluation is proportionate and aligned to claims.
  • PMS and PMCF demonstrate control over real-world performance.

The NB review process accelerates when the documentation structure allows rapid verification of consistency. It slows down when the reviewer has to reconcile contradictions between sections.

Step 3. Clinical evidence strategy for dental devices: CER, PMCF, and clinical investigations.

Clinical evidence is typically the largest schedule and cost driver under MDR. The core issue is not volume; it is relevance and alignment to the intended purpose, claims, and risk profile.

A manufacturer-ready clinical evidence strategy generally combines:

  • A Clinical Evaluation Report (CER) anchored to intended purpose and claims.
  • Literature appraisal aligned with state of the art and device technology.
  • Post-market surveillance (PMS) data, including complaint trending and failure mode monitoring.
  • A PMCF approach proportionate to risk and uncertainty, designed to confirm performance and safety in clinical use.

The webinar highlighted a recurring decision point for manufacturers: when “well-established technology” is an appropriate foundation for clinical justification, and when novelty requires additional clinical evidence.

Practical examples discussed in dental context:

  • Implantables may be approached differently depending on the maturity of the technology, materials, and the presence of novel surface engineering or expanded clinical claims.
  • Resorbable grafting materials and membranes often require particular attention to degradation behaviour, degradation by-products, biological response, and performance endpoints over time.
  • Legacy devices require a structured assessment of whether existing data remains fit for MDR expectations and current clinical practice.

Notified Bodies evaluate clinical evidence against the totality of the file. Evidence is rarely accepted in isolation if risk management, claims, and verification logic do not support it.

How to pass a Notified Body Assessment as a dental device manufacturer

In practice, manufacturers do not lose time because they miss a single document. They lose time when their file is not reviewable as a system. Notified Bodies move more efficiently when traceability is clear and when the submission reads as one coherent position.

A practical Assessment-readiness checklist:

  • Intended purpose and claims are consistent across IFU, labeling, CER, and technical documentation.
  • Classification is justified with explicit Annex VIII logic and references.
  • Risk management is device-specific and connects hazards to controls and to verification evidence.
  • Verification and validation are planned around worst cases, critical characteristics, and real-use conditions.
  • Clinical evaluation is aligned to claims and supported by appropriate PMS/PMCF.
  • PMS/PMCF plans are measurable, realistic, and proportionate.

The NB-ready evidence pack (manufacturer checklist)

For internal alignment and faster NB interaction, a compact evidence pack is often effective. A practical NB-ready pack includes:

  • Device description, intended purpose, and claim list (single source of truth)
  • Annex VIII classification memorandum (rule logic and rationale)
  • GSPR checklist with clear evidence references
  • Risk management file (ISO 14971) and traceability to controls
  • Verification and validation plan and reports (including worst-case rationale)
  • Biological evaluation strategy (ISO 10993) and, where relevant, extractables and leachables considerations
  • Mechanical performance evidence, acceptance criteria, and justification of test conditions
  • Clinical evaluation (CER), literature strategy, and equivalence rationale when applicable
  • PMS plan and PMCF plan/report (tailored to failure modes and risk profile)
  • Labeling and IFU, including UDI readiness and EUDAMED considerations as applicable

The evidence pack is not a marketing summary. It is a reviewer tool. Its value is to reduce time spent locating and reconciling information across the file.

How to categorise dental devices under EU MDR (classification shortcuts by product type)

Dental manufacturers rarely struggle with the existence of Annex VIII. The difficulty is applying it consistently across a portfolio that mixes implantables, materials, digital workflows, and patient-specific outputs. A practical way to reduce ambiguity is to group dental devices by product type first, then confirm the applicable MDR rules and evidence expectations.

  • Dental implants and abutments
    Typical focus areas include the Annex VIII rationale (often centred on implantability), the scope of compatibility claims (system-specific definition), mechanical performance evidence, and the regulatory impact of surface treatments and coatings.
  • Resorbable grafting materials, membranes, and hemostats
    These products often require careful justification of degradation behaviour, degradation by-products, biological response over time, and clinically relevant endpoints. Evidence expectations frequently increase with resorption and biological interaction.
  • Restoratives and CAD/CAM materials (including semi-finished products)
    Categorisation often depends on intended purpose and how the manufacturer positions the product (medical device versus material). Performance claims, manufacturing controls, and labeling language typically drive both classification stability and clinical evidence requirements.
  • Custom-made versus patient-matched devices
    The practical distinction is design control. Custom-made devices require a prescription-led pathway with specific documentation, while patient-matched outputs often operate within a validated design envelope controlled by the manufacturer. This distinction materially affects technical documentation and lifecycle obligations.
  • Dental software, AI, and digital workflows
    Software categorisation under Rule 11 depends on how the software influences clinical decisions and outcomes. Validation, cybersecurity, and change control become central, and AI governance requirements can intersect with technical documentation expectations.

Common dental device categories to review:

A structured categorisation approach helps manufacturers align intended purpose, classification rationale, verification planning, and clinical evidence strategy across the full file, which reduces clarification cycles during review.

To learn more about dental device regulation and compliance support, read: Regulatory Compliance for Dental Products.

PMS, PMCF, UDI/EUDAMED, and legacy devices: maintaining market access

Under MDR, post-market obligations are part of the evidence lifecycle. PMS and PMCF support the ongoing demonstration of safety and performance and can become decisive during renewals and significant changes.

Practical PMS and PMCF considerations in dental:

  • Define and monitor meaningful signals: failures, revisions, fractures, loosening, complaint patterns, and trends linked to known failure modes.
  • Ensure PMCF is proportionate and focused on remaining uncertainty, rather than generic data collection.
  • Maintain a controlled approach to labeling updates, complaint handling, and vigilance reporting.

For legacy devices, manufacturers should perform a structured assessment of whether existing clinical and post-market data remains sufficient for current claims, current clinical practice, and MDR expectations. Where gaps exist, a targeted plan that links risk management to clinical evaluation and post-market follow-up reduces uncertainty and supports continuity of market access.

MDR transition in practice: Argen dental alloys case study

Many MDR transitions fail for predictable reasons: legacy documentation does not meet MDR expectations, clinical evaluation lacks alignment to intended purpose and risk management, and post-market systems are not mature enough to support lifecycle obligations. A recent example from dental materials demonstrates what a structured transition can look like in practice.

Argen transitioned a legacy dental alloy portfolio to Regulation (EU) 2017/745 for use in fabricating full-cast and ceramic-veneered restorations (including crowns, bridges, and removable partial dentures). The project required closing historical Notified Body non-conformities and strengthening core regulatory processes to achieve an audit-ready MDR position

What the work focused on:

  • Strengthening the clinical evaluation framework
  • Structuring the PMS system
  • Aligning risk management with MDR and ISO 14971
  • Establishing MDR-compliant biological evaluation
  • Aligning essential technical documentation

To read the full case study, click here: Case Study: How MDx Enabled Argen’s Successful Transition to MDR CE Marking for Dental Alloys.

Frequently Asked Questions About Dental Device Regulation

Do legacy (MDD) dental devices require new clinical data under EU MDR 2017/745?

Not automatically. The decision depends on risk profile, intended purpose, claims, and whether the technology can be justified as well established using appropriate literature, PMS data, and alignment with the state of the art.

Custom-made vs patient-matched in dentistry: what is the practical difference, and what does it change?

The practical difference is design control. Custom-made devices rely on a written prescription and patient-specific specifications defined by the healthcare professional. Patient-matched devices are typically produced within a validated “design envelope” controlled by the manufacturer. This distinction affects documentation expectations, validation logic, and lifecycle obligations.

What validation evidence is expected for patient-matched dental devices?

Manufacturers should define a validated design envelope, identify worst-case configurations within that envelope, and validate the worst case in its final condition. This approach supports a defensible argument that the full envelope remains safe and performs as intended.

Why do resorbable grafts and membranes face higher scrutiny, and what evidence is usually questioned?

Because risk and performance evolve over time as the device degrades. Evidence discussions often focus on degradation behaviour, degradation by-products, biological response, and clinically relevant performance endpoints across the functional period.

When does dental software fall under MDR Rule 11, and what is the fastest way to reduce Notified Body back-and-forth?

When software influences clinical decisions or outcomes, Rule 11 can apply and classification may increase, triggering stronger validation and control expectations. To reduce review cycles, manufacturers should maintain consistency across intended purpose and claims, classification rationale, risk management, verification evidence, clinical evaluation, and PMS/PMCF. Inconsistencies are a major driver of clarification rounds and timeline slippage

Written by:
Alberto Bardají

Alberto Bardají

Head of Medical Devices

Senior med-tech expert & ex-Notified Body reviewer with deep experience in high-risk implants, orthopedics, dental & neurology.